Dalotech - Data Logger information

Part Number:  ULTRASHOCK-EB-50
Price:   $899.00

MadgeTech UltraShock-EB

The MadgeTech UltraShock-EB is a stand alone, battery operated, temperature, humidity, pressure and 3-axis shock data recorder. Continuously sampling at 500Hz, the shock sensors record time and date stamped peak acceleration data at the user-defined intervals that temperature, humidity and pressure readings are stored. Shock data may be displayed for the three axis as well as a calculated vector sum to provide overall acceleration/shock information, along with the temperature, humidity and pressure data. MadgeTech softwarealso provides a summary of statistics for average, minimum and maximum values for all of the sensors. Available in three ranges(±5g, ±50g, ±100g), with extended battery life. A software and cable kit is required for operation.

Temperature Specifications
Sensor: Semiconductor
Range: -20 to +54ºC
Resolution: 0.1ºC
Accuracy: ±0.5ºC (0 to +50ºC)
 
Humidity Specifications
Sensor: Capacitive Polymer
Range: 0 to 95%RH
Resolution: 0.1%RH
Calibrated Accuracy: ±3%RH (±2%RH typical at 25ºC)
Specified Accuracy Range: +10 to +40°C; 10 to 80%RH
 
Pressure Specifications
Sensor: Semiconductor Strain Gage
Range: 0 to 30PSIA
Resolution: 0.002PSIA
Calibrated Accuracy: ±1.0%FSR at 25ºC; ±0.2% typical
Specified Accuracy Range: 0 to 30 PSIA, 25°C
 
Shock Specifications
Accelerometer Type: MEMS Semiconductor
Absolute Max. Acceleration: 200g (all versions)
Sampling Rate: 1.953ms/512Hz
Accelerometer Freq. Resp.: 0Hz to approx. 400Hz
UltraShock-EB Ranges
Acceleration Range: ±5 g ±50 g ±100 g
Acceleration Accuracy: ±0.2 g ±1 g ±2 g
Acceleration Resolution: 0.01 g 0.05 g 0.1 g
General Specifications
Readings: 174,762 per channel (1,572,858 total readings)
Recording Interval: 64Hz to 5 minutes for shock, selectable in software. Temperature, pressure and humidity, approximately every 2 seconds at intervals faster than 2 seconds, otherwise sampled at the reading rate.
Start Modes: Immediate start or delay start, up to 180 days from PC launch
Password Protection: An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password.
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Power: 6 D-cell alkaline batteries, user replaceable
Data Format: Date and time stamped gravities (g and mg), temperature (°C, °F, K, °R), humidity (%RH, mg/ml water vapor concentration), pressure (PSIA, inHg, mmHg, bar, atm, Torr, Pa, kPa, MPa)
Time Accuracy: ±1 minute/month at 20 to 30°C
User Replaceable Battery: 60 days typical at 25°C, 1 minute reading rate
Computer Interface: PC serial or USB (interface cable required); 115,200 baud
Computer Interface: PC serial or USB (interface cable required); 115,200 baud
Software: Windows 95/98/ME/NT/2000/XP based software
Operating Environment: -20 to +54ºC, 0 to 95%RH (non-condensing)
Operating Environment: -20 to +54ºC, 0 to 95%RH (non-condensing)
Dimensions: 5.5” x 5.4” x 3.2” (140mm x 137mm x 80mm)
Weight: 80 oz. (2.3kg)
Enclosure: Anodized aluminum
Approvals: CE
Ordering Information
View Shopping Cart
Description Part Number Qty 1-9 Qty 10+  
 
Data Logger
MadgeTech UltrashockTemperature, Humidity, Pressure, and Shock Recorder, -50 to +50G, Extended Battery ULTRASHOCK-EB-50 $899.00 contact us Contact us to order
MadgeTech UltrashockTemperature, Humidity, Pressure, and Shock Recorder, -5 to +5G, Extended Battery ULTRASHOCK-EB-5 $899.00 contact us Contact us to order
MadgeTech UltrashockTemperature, Humidity, Pressure, and Shock Recorder, -100 to +100G, Extended Battery ULTRASHOCK-EB-100 $899.00 contact us Contact us to order
 
Software and Communication Cable (One required)
MadgeTech Software and RS-232 Interface Cable Kit IFC110 $99.00 contact us Contact us to order
MadgeTech Software and USB Interface Cable Kit IFC200 $119.00 contact us Contact us to order



© 2018 Dalotech LLC